Comparison · NCT01665053-sae
Promus Element Plus vs SYNERGY: serious adverse events
Calculated by ConductScience from the counts the trial posted to ClinicalTrials.gov.
2 sourcestrial-comparisons-2026-03-23-0.1.0
Cite this record
ConductScience Library. Promus Element Plus vs SYNERGY: serious adverse events. Trial comparisons, record NCT01665053-sae. Release trial-comparisons-2026-03-23-0.1.0, 23 Mar 2026. https://conductscience.org/library/comparisons/NCT01665053-sae
On this page
Measures6
Calculated by ConductScience from the counts the trial posted to ClinicalTrials.gov.
| Measure | Value | 95% interval, lower | 95% interval, upper | Method | Note |
|---|---|---|---|---|---|
| Risk in group 1 | 0.313 | 0.282 | 0.345 | wilson | |
| Risk in group 2 | 0.313 | 0.283 | 0.345 | wilson | |
| Risk difference, group 1 minus group 2 | -0.00059 | -0.0448 | 0.0436 | newcombe hybrid score | |
| Risk ratio, group 1 over group 2 | 0.998 | 0.866 | 1.15 | log | |
| Odds ratio, group 1 over group 2 | 0.997 | 0.812 | 1.23 | log | |
| Number needed to treat, for one fewer participant with a serious adverse event on Promus Element Plus than on SYNERGY | 1,700 | from the risk difference interval | the interval includes no difference |
Source ClinicalTrials.gov
Groups2
Neither group, or both, names a comparator, so the groups keep ClinicalTrials.gov's order.
| Group | Title | Description | With the outcome | Analysed | ClinicalTrials.gov code |
|---|---|---|---|---|---|
| Group 1 | Promus Element Plus | PROMUS Element Plus is a device/drug combination product composed of two components, a device (coronary stent system including a platinum chromium stent platform) and a drug product (a formulation of everolimus contained in a polymer coating).~PROMUS Element Plus: A drug eluting coronary stent system | 262 | 838 | EG000 |
| Group 2 | SYNERGY | SYNERGY is a device/drug combination product composed of two components, a device (coronary stent system including a platinum chromium stent platform) and a drug product (a formulation of everolimus contained in a bioabsorbable polymer coating).~SYNERGY: A drug eluting coronary stent system | 265 | 846 | EG001 |
Source ClinicalTrials.gov
Harm
- What it counts
- Total, serious adverse events
- Time frame
- Site-Reported Serious Adverse Events to 12 Months
Source ClinicalTrials.gov
Trial
- Title
- The EVOLVE II Clinical Trial To Assess the SYNERGY Stent System for the Treatment of Atherosclerotic Lesion(s)
- NCT number
- NCT01665053
- Phases
- PHASE3
- Conditions
- Coronary Artery Disease
- Interventions
- PROMUS Element Plus, SYNERGY
- Lead sponsor
- Boston Scientific Corporation
- Results first posted
- 2016-01-07
- Results on ClinicalTrials.gov
- https://clinicaltrials.gov/study/NCT01665053?tab=results
Source ClinicalTrials.gov
Source
- Release
- trial-comparisons-2026-03-23-0.1.0
- Release manifest SHA-256
- f66225897fbf258b57ea10bdacfd0347d3cdfda93f7f5685234b7e06ea36fa57
- AACT copy
- 2026-03-23
- Tables read
- outcome_counts.parquet, outcome_measurements.parquet, outcomes.parquet, reported_event_totals.parquet, reported_events.parquet, result_groups.parquet, studies.parquet, ctgov_trial.parquet
Source ClinicalTrials.gov
Sources 2
- ClinicalTrials.gov's posted results, through CTTI's AACTFreely available. Acknowledgement requested
- ClinicalTrials.govNo license stated
Each source's terms
- ClinicalTrials.gov's posted results, through CTTI's AACTClinicalTrials.gov data via Aggregate Analysis of ClinicalTrials.gov (AACT), CTTI. CTTI makes its materials freely available and requests acknowledgement. https://aact.ctti-clinicaltrials.org/documentation/299
- ClinicalTrials.govnone stated
ClinicalTrials.gov's data carry an international copyright outside the United States: https://clinicaltrials.gov/about-site/terms-conditions. The measures are ConductScience's calculation from them.
For software
Read this comparison from software.
The record is JSON. No key.