ConductScience.org, home

Comparison · NCT01751906-028e5e39

Absorb BVS vs XIENCE: Number of Cardiac Death/All MI

Calculated by ConductScience from the counts the trial posted to ClinicalTrials.gov.

2 sourcestrial-comparisons-2026-03-23-0.1.0

Download JSON
Cite this record

ConductScience Library. Absorb BVS vs XIENCE: Number of Cardiac Death/All MI. Trial comparisons, record NCT01751906-028e5e39. Release trial-comparisons-2026-03-23-0.1.0, 23 Mar 2026. https://conductscience.org/library/comparisons/NCT01751906-028e5e39

On this page

Measures6

Calculated by ConductScience from the counts the trial posted to ClinicalTrials.gov.

MeasureValue95% interval, lower95% interval, upperMethodNote
Risk in group 10.1560.1360.178wilson
Risk in group 20.1320.1080.162wilson
Risk difference, group 1 minus group 20.0235-0.01160.0563newcombe hybrid score
Risk ratio, group 1 over group 21.180.9241.5log
Odds ratio, group 1 over group 21.210.9131.6log
Number needed to treat, for one more participant with the outcome on Absorb BVS than on XIENCE42.5from the risk difference intervalthe interval includes no difference

Groups2

Neither group, or both, names a comparator, so the groups keep ClinicalTrials.gov's order.

GroupTitleDescriptionWith the outcomeAnalysedClinicalTrials.gov code
Group 1Absorb BVS
Subjects receiving Absorb BVS~Absorb BVS: • Scaffold diameters: 2.5, 3.0 and 3.5 mm~* Scaffold lengths: 8, 12, 18, and 28 mm The 3.0 x 18 mm Absorb BVS will be used for the Lead-In. Both the 8 mm and 12 mm lengths will be available for the 2.5/3.0 mm diameter Absorb BVS. Only the 12 mm length will be available for the 3.5 mm diameter.~* The commercially approved CE marked device will be used in geographies where it is commercially available. The commercially approved CE marked 23mm Absorb BVS device will not be used in this study.~Bioabsorbable drug eluting stent implantation for improving coronary luminal diameter in patients, including those with diabetes mellitus, with ischemic heart disease due to de novo native coronary artery lesions (length ≤ 24 mm) with a reference vessel diameter of ≥ 2.5 mm and ≤ 3.75 mm.
1851,187OG000
Group 2XIENCE
Subjects receiving XIENCE V, XIENCE PRIME, or XIENCE Xpedition~XIENCE: Commercially approved XIENCE Family Stent System, inclusive of XIENCE V, XIENCE PRIME, XIENCE Xpedition, XIENCE Alpine, XIENCE Pro (OUS only), and XIENCE ProX (OUS only).~* Stent diameters: 2.5, 2.75, 3.0, 3.25, 3.5 and 4.0 mm~* Stent lengths: 8, 12, 15, 18, 23, and 28 mm. The 3.25 mm is only available for XIENCE Xpedition~* For geographies where these devices are commercially available, the investigational sties may use only their locally approved devices~To improve coronary luminal diameter in patients, including those with diabetes mellitus, with ischemic heart disease due to de novo native coronary artery lesions (length ≤ 24 mm) with a reference vessel diameter of ≥ 2.5 mm and ≤ 3.75 mm.
81612OG001

Outcome

Type
SECONDARY
Title
Number of Cardiac Death/All MI
Description
All deaths includes Cardiac death: Any death due to proximate cardiac cause (e.g. MI, low-output failure, fatal arrhythmia), unwitnessed death and death of unknown cause, all procedure related deaths including those related to concomitant treatment.~Vascular death: Death due to non-coronary vascular causes such as cerebrovascular disease, pulmonary embolism, ruptured aortic aneurysm, dissecting aneurysm, or other vascular cause.~Non-cardiovascular death: Any death not covered by the above definitions such as death caused by infection, malignancy, sepsis, pulmonary causes, accident, suicide or trauma.~Myocardial Infarction (MI) - Q wave MI: Development of new, pathological Q wave on the ECG.~-Non-Q wave MI: Those MIs which are not Q-wave MI
Time frame
0 to 5 years
Population
ITT set. The number of participants analyzed includes subjects who had available follow up data at that time frame.

Source ClinicalTrials.gov

Trial

Title
ABSORB III Randomized Controlled Trial (RCT)
NCT number
NCT01751906
Phases
NA
Conditions
Coronary Artery Disease, Coronary Artery Stenosis, Coronary Disease, Coronary Stenosis
Interventions
Absorb BVS, XIENCE
Lead sponsor
Abbott Medical Devices
Results first posted
2018-02-13

Source ClinicalTrials.gov

Source

Release
trial-comparisons-2026-03-23-0.1.0
Release manifest SHA-256
f66225897fbf258b57ea10bdacfd0347d3cdfda93f7f5685234b7e06ea36fa57
AACT copy
2026-03-23
Tables read
outcome_counts.parquet, outcome_measurements.parquet, outcomes.parquet, reported_event_totals.parquet, reported_events.parquet, result_groups.parquet, studies.parquet, ctgov_trial.parquet

Source ClinicalTrials.gov

Sources 2

  • ClinicalTrials.gov's posted results, through CTTI's AACTFreely available. Acknowledgement requested
  • ClinicalTrials.govNo license stated
Each source's terms
  • ClinicalTrials.gov's posted results, through CTTI's AACTClinicalTrials.gov data via Aggregate Analysis of ClinicalTrials.gov (AACT), CTTI. CTTI makes its materials freely available and requests acknowledgement. https://aact.ctti-clinicaltrials.org/documentation/299
  • ClinicalTrials.govnone stated

ClinicalTrials.gov's data carry an international copyright outside the United States: https://clinicaltrials.gov/about-site/terms-conditions. The measures are ConductScience's calculation from them.

For software

Read this comparison from software.

The record is JSON. No key.

GET https://conductscience.org/api/comparisons/v1/NCT01751906-028e5e39