ConductScience.org, home

Comparison · NCT03168776-sae

BuMA Supreme Coronary Stent System (ITT) vs Durable Polymer EES (XIENCE or Promus DES) (ITT): serious adverse events

Calculated by ConductScience from the counts the trial posted to ClinicalTrials.gov.

2 sourcestrial-comparisons-2026-03-23-0.1.0

Download JSON
Cite this record

ConductScience Library. BuMA Supreme Coronary Stent System (ITT) vs Durable Polymer EES (XIENCE or Promus DES) (ITT): serious adverse events. Trial comparisons, record NCT03168776-sae. Release trial-comparisons-2026-03-23-0.1.0, 23 Mar 2026. https://conductscience.org/library/comparisons/NCT03168776-sae

On this page

Measures6

Calculated by ConductScience from the counts the trial posted to ClinicalTrials.gov.

MeasureValue95% interval, lower95% interval, upperMethodNote
Risk in group 10.1930.1710.218wilson
Risk in group 20.2080.1760.244wilson
Risk difference, group 1 minus group 2-0.0147-0.05720.0256newcombe hybrid score
Risk ratio, group 1 over group 20.9290.7581.14log
Odds ratio, group 1 over group 20.9120.7061.18log
Number needed to treat, for one fewer participant with a serious adverse event on BuMA Supreme Coronary Stent System (ITT) than on Durable Polymer EES (XIENCE or Promus DES) (ITT)67.9from the risk difference intervalthe interval includes no difference

Groups2

Neither group, or both, names a comparator, so the groups keep ClinicalTrials.gov's order.

GroupTitleDescriptionWith the outcomeAnalysedClinicalTrials.gov code
Group 1BuMA Supreme Coronary Stent System (ITT)
The BuMA Supreme Biodegradable Drug Coated Coronary Stent System is a device / drug combination product consisting of a drug-coated balloon expandable stent and a rapid exchange delivery system. The cobalt chromium (CoCr) stent is coated with a very thin, non-erodable layer of poly n-butyl methacrylate (PBMA) that is covalently bonded to the metal surface through a proprietary electro-grafting (eG™) process. A topcoat is then applied that contains sirolimus, the active ingredient, embedded in a biodegradable polymer, poly lactic co-glycolic acid (PLGA).
2101,086EG000
Group 2Durable Polymer EES (XIENCE or Promus DES) (ITT)
The control comparator is durable polymer everolimus-eluting stent systems (DP EES), consisting of the Xience™ family of durable polymer cobalt chromium everolimus-eluting coronary stent systems (Abbott Vascular, Santa Clara, CA, US), and the PROMUS Element™ family of everolimus-eluting coronary stent systems.
113543EG001

Harm

What it counts
Total, serious adverse events
Time frame
5 Years

Source ClinicalTrials.gov

Trial

Title
PIONEER III Trial to Assess Safety and Efficacy of the BuMA Supreme™ Drug Coated Coronary Stent in Patients With Coronary Disease
NCT number
NCT03168776
Phases
NA
Conditions
Coronary Artery Disease
Interventions
BuMA Supreme DES, Xience or Promus DES
Lead sponsor
Sino Medical Sciences Technology Inc.
Results first posted
2026-01-27

Source ClinicalTrials.gov

Source

Release
trial-comparisons-2026-03-23-0.1.0
Release manifest SHA-256
f66225897fbf258b57ea10bdacfd0347d3cdfda93f7f5685234b7e06ea36fa57
AACT copy
2026-03-23
Tables read
outcome_counts.parquet, outcome_measurements.parquet, outcomes.parquet, reported_event_totals.parquet, reported_events.parquet, result_groups.parquet, studies.parquet, ctgov_trial.parquet

Source ClinicalTrials.gov

Sources 2

  • ClinicalTrials.gov's posted results, through CTTI's AACTFreely available. Acknowledgement requested
  • ClinicalTrials.govNo license stated
Each source's terms
  • ClinicalTrials.gov's posted results, through CTTI's AACTClinicalTrials.gov data via Aggregate Analysis of ClinicalTrials.gov (AACT), CTTI. CTTI makes its materials freely available and requests acknowledgement. https://aact.ctti-clinicaltrials.org/documentation/299
  • ClinicalTrials.govnone stated